Pelvic health devices
A handheld therapy device that opens a fraction of a millimeter at a time, and a childbirth device that knows when to stop pushing.
- Years
- 2014–2019
- Boards
- Every revision, prototype to production
- Firmware
- All of it: device, charging case, trial units
- Disciplines
- Electronics, Firmware, Medical devices, Manufacturing
A therapy device you hold in one hand
The first device treats pelvic pain by opening gently, a set step at a time, to a size the patient chooses on two buttons and reads on a small display. It vibrates, charges in its case, and everything has to fit inside its handle.
The board in the drawing is its controller, shaped to fit that handle: a Bluetooth system-on-chip, boost converters and regulators for a single lithium cell, a motor drive and an optical encoder that counts how far the mechanism has turned.
Firmware that knows where it is
The firmware tracks the mechanism through its encoder, homes against a limit switch, turns position into diameter through a calibration table, and drives the motor, vibrator, buzzer and display. It also charges the battery itself, constant current then constant voltage, and shuts down safely when the cell runs low.
The charging case got its own small board and firmware, written in assembly for a tiny microcontroller, to tell the user when the device is charging and when it's done.
From the bench to production
We built the early units ourselves: reflowed the boards, assembled the devices and handed them to the team, including the batch for their first trial. As the design settled we took on cost reduction, revised the boards for the contract manufacturer, and wrote the programming scripts and specifications their line used to load and test every unit. Later, a bench logger drove a programmable supply to charge devices and record each one by serial number, to prove the batteries.
A childbirth device with a sense of force
The second device is used during labor. Its arms open to a set diameter, and it must never push harder than is safe. We designed its main board around a load cell read by a 16-bit converter, an optical encoder and an Arm microcontroller, and wrote its trial firmware: a state machine for its buttons, a force limit that changes with diameter, and a recorder that logs position, force and every event, second by second, for review after each use.
What we did
- Schematics and PCB layout for every board: the controllers, a limit-switch board, the charging case and test adapters, through more than a dozen revisions.
- All the firmware: motor and encoder control, calibration, user interface, battery charging and the charging case.
- Wireless charging bring-up, power design and low-power shutdown.
- Prototype builds, production programming and battery test tooling.
- The main board and trial firmware for the childbirth device, with force limiting and data logging.